The Food and Drug Administration announced on Aug. 20, 2026, that Bard Peripheral Vascular Inc is recalling its EnCor EnCompass Breast Biopsy and Tissue Removal System Console and Driver due to potential calibration errors. These defects may interrupt biopsy procedures or necessitate additional tissue sampling.
How to tell if you have it
- Brand: Bard Peripheral Vascular
- Product: EnCor EnCompass Breast Biopsy and Tissue Removal System Console (model number: ENCPCONSOLE) and Driver (model number: ENCPDRIVER)
- Software Versions: 1.1.1.10207 and 1.2.0.12133
- UDI-DI: 00801741235221 and 00801741234248
- Serial Numbers: VTLNPX06, VTLNPX0F, VTLNPX0D, VTLNPX08, VTLNPX0K, VTLNPX0A, VTLNPX0C, VTLNPX0L, VTLNPX0J, VTLNPX04, VTLNPX0H, VTLNPX02, VTLPPX1G, VTLPPX1D, VTLPPX1C, VTLPPX1M, VTLPPX2H, VTLPPX2I, VTLPPX2F, VTLQPX15, VTLQPX1Q, VTLQPX1G, VTLQPX14, VTLQPX0T, VTLQPX1Y, VTLQPX1M, VTLQPX2G, VTLQPX1O, VTLQPX1K, VTLQPX19, VTLQPX2E, VTLQPX1W, VTLQPX10, VTLQPX1Z, VTLQPX24, VTLQPX21, VTLQPX0X, VTLQPX1E, VTLQPX0U, VTLQPX26, VTLQPX2F, VTLQPX25, VTLQPX0Z, VTLQPX18, VTLQPX28, VTLQPX20, VTLQPX27, VTLQPX11, VTLQPX2B, VTLQPX1F, VTLQPX2H, VTLQPX41, VTLQPX3T, VTLQPX43, VLRPX0R, VTKYPX09, VTKYPX44, VTKYPX46, VTKYPX47, VTKYPX48, VTKYPX49, VTKZPX14, VTKZPX15, VTKZPX20, VTKZPX17, VTLNPX4I, VTLNPX49, VTLNPX4G, VTLNPX4N, VTLNPX4E, VTLNPX4L, VTLNPX44, VTLNPX46, VTLNPX4C, VTLPPX27, VTLPPX28, VTLPPX29, VTLPPX1Z, VTLPPX37, VTLPPX3B, VTLPPX34, VTLPPX38, VTLPPX36, VTLQPX0Q, VTLQPX0B, VTLQPX0E, VTLQPX0P, VTLQPX0F, VLRPX0X, VLRPX0Z, VTKYPX07, VTKYPX06, VTKYPX04, VTLNPX4B, VTLNPX42, VTLNPX4F, VTLPPX3D, VTLQPX0M, VTKYPX51
- Units affected: 99
The hazard
According to the FDA, the system may experience probe calibration or handheld driver failure (code HH01-00020009) and pressure sensor errors (code E01-0054000E). These errors prompt users to restart and re-calibrate, which may lead to interruption or prolongation of the biopsy procedure, or the need for additional tissue sampling or repeat biopsy.
Context
GovInfoNow's records show Bard Peripheral Vascular Inc has issued 40 other recalls since Sept. 22, 2017. Other recent recalls include a Dec. 12, 2025, recall for Broviac Cutdown Trays due to sterile barrier compromise risk and an Aug. 21, 2025, recall for Venclose digiRF Generators that may cause false failures. Other recent recalls of similar products include a May 21, 2026, Bard Soft Tissue Biopsy Trays Recall for lidocaine ampoule manufacturing deficiency and an April 24, 2026, HydroMARK Breast Biopsy Marker Recall for incorrect labeling risk.