Emano Metrics recalls v2.0.0 software for Emano Flow devices due to inaccurate urine flow measurements. The software version may over-report flow rates and volumes, affecting medical assessments.
Problem description/Reason for recall: During the period March 22, 2026, to May 29, 2026, recorded urination audio was captured at 48,000 Hz while the model was trained and validated on 44,100 Hz audio. The models depend on the sampling so every recording in the window was therefore processed against a time/frequency scale the models did not expect, producing systematically biased Qmax, Qmean, and voided-volume estimates. The problem usually over-reports flow rate and volume; the firm have made corrected results available to care teams.
Emano Metrics recalls Emano Flow devices with software version v2.0.0 because they inaccurately measure urine flow rates and volumes. This could lead to incorrect medical assessments if used for health monitoring.
The software version v2.0.0 processes urination audio at 48,000 Hz instead of the 44,100 Hz standard used for training. This causes systematic errors in flow rate and volume calculations, leading to over-reported measurements.
Users of Emano Flow devices with software version v2.0.0, particularly those using the device for medical urination flow monitoring. Individuals with specific medical conditions requiring precise flow measurements may be at higher risk.
Check if your Emano Flow device has software version v2.0.0. The product is identified by GS1 GTIN code 199874903484 and is distributed via mobile application.
Emano Metrics has provided corrected results for accurate measurements. Check for software updates through the mobile application.
No, the device is safe to use but the software version v2.0.0 may provide inaccurate measurements. Emano Metrics has corrected the issue and offers updated software.
Check your Emano Flow device for software version v2.0.0. The product is identified by GS1 GTIN code 199874903484 and is distributed via mobile application.
Update to the latest software version through the mobile application to ensure accurate measurements. Emano Metrics provides corrected results for medical care teams.
We’ve drafted a message you can send Emano Metrics to request your refund or repair — edit it as you like.
Subject: Recall Z-2777-2026 — Emano Metrics Emano Metrics Hello, I own a Emano Metrics that is covered by recall Z-2777-2026 from Emano Metrics. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.