Skip to content
Lower risk Medical device recall · FDA · Published June 8, 2026

Micro Therapeutics Pipeline Flex Embolization Device: Incorrect Labeling Risk

Micro Therapeutics recalls Pipeline Flex Embolization Devices with incorrect labeling that may cause improper use in intracranial aneurysm treatment.

🇺🇸 FDA recall #Z-2774-2026Medical deviceMicro Therapeutics ›
✎ GovInfoNow plain-language explainer

Micro Therapeutics recalls certain Pipeline Flex Embolization Devices due to incorrect labeling. Devices labeled as 3.0 mm x 20 mm implants actually contain 12 mm implants, and those labeled as 3.0 mm x 12 mm implants contain 20 mm implants. This could lead to improper treatment if used without correction. Read more at govinfonow.com/recall/127822-micro-therapeutics-pipeline-flex-embolization-device-incorrect-labeling-risk.

Source Summary by GovInfoNow · govinfonow.com/recall/127822-micro-therapeutics-pipeline-flex-embolization-device-incorrect-labeling-risk · Updated August 18, 2026 · FDA enforcement record: FDA #Z-2774-2026

Cite this page

“Micro Therapeutics Pipeline Flex Embolization Device: Incorrect Labeling Risk.” GovInfoNow, updated August 18, 2026. https://govinfonow.com/recall/127822-micro-therapeutics-pipeline-flex-embolization-device-incorrect-labeling-risk. Based on FDA recall #Z-2774-2026 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2774-2026%22&limit=1).

The FDA announced on June 8, 2026, that Micro Therapeutics, Inc. is recalling the Pipeline Flex Embolization Device with Shield Technology due to incorrect labeling hazards. The recall involves devices where the labeled dimensions do not match the actual implant size.

How to tell if you have it

  • Brand: Micro Therapeutics
  • Product: Pipeline Flex Embolization Device with Shield Technology, 3.00 mm x 20 mm 027 REF PED2-300-20
  • UDI: 00763000448844
  • Lot: D062531
  • Quantity: 6 units

The hazard

According to the FDA, embolization devices labeled as 3.0 mm x 20 mm implants contain 12 mm implants, and devices labeled as 3.0 mm x 12 mm implants contain 20 mm implants.

Context

GovInfoNow's records show Micro Therapeutics, Inc. has 6 other recalls on record since July 13, 2021. Other recent recalls of similar products include EMPOWR 3D KNEE Implants Have Incorrect Labeling (Jan. 15, 2026), Centinel Prodisc C SK U.S. Implant Extra Large 5mm - Labeling Mixup Recall (March 20, 2026), and W L Gore GORE VIABIL Short Wire Biliary Endoprosthesis Recall: Incorrect Labeling Risk (May 22, 2026).

How to tell if you have this one

Check for the product name 'Pipeline Flex Embolization Device with Shield Technology' and the model number 3.00 mm x 20 mm 027 REF PED2-300-20. The device was used for intracranial aneurysm treatment.

  • Brand
    Micro Therapeutics
  • Category
    Medical device
FDA recall event · 2 products

All 2 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Ongoing
  • Distribution
    US Nationwide distribution in the states of CT, DE, IN, MA, NC, NJ, NM, NY, OH, PA, TX, VA, WA.
Recall numberProduct, codes & lotsQuantity
Z-2773-2026Pipeline Flex Embolization Device with Shield Technology, 3.00 mm x 12 mm 027 REF PED2-300-12. Intracranial aneurysm flow diverter; UDI: 00763000448806/ Lot: D06261211 units
Z-2774-2026Pipeline Flex Embolization Device with Shield Technology, 3.00 mm x 20 mm 027 REF PED2-300-20. Intracranial aneurysm flow diverter; UDI: 00763000448844/ Lot: D0625316 units
Based on the official FDA remedy

What to do — step by step

  1. Verify Device Label

    Check the device label for the model number and size specifications to confirm if it matches the recalled product.

Where it went and when

Where it was distributed

Distributed nationwide; the notice names 13 states.

Alaska: nationwideAKMaine: nationwideMEVermont: nationwideVTNew Hampshire: nationwideNHWashington: named in the noticeWAIdaho: nationwideIDMontana: nationwideMTNorth Dakota: nationwideNDMinnesota: nationwideMNIllinois: nationwideILWisconsin: nationwideWIMichigan: nationwideMINew York: named in the noticeNYRhode Island: nationwideRIMassachusetts: named in the noticeMAOregon: nationwideORNevada: nationwideNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: nationwideIAIndiana: named in the noticeINOhio: named in the noticeOHPennsylvania: named in the noticePANew Jersey: named in the noticeNJConnecticut: named in the noticeCTCalifornia: nationwideCAUtah: nationwideUTColorado: nationwideCONebraska: nationwideNEMissouri: nationwideMOKentucky: nationwideKYWest Virginia: nationwideWVVirginia: named in the noticeVAMaryland: nationwideMDDelaware: named in the noticeDEArizona: nationwideAZNew Mexico: named in the noticeNMKansas: nationwideKSArkansas: nationwideARTennessee: nationwideTNNorth Carolina: named in the noticeNCSouth Carolina: nationwideSCDistrict of Columbia: nationwideDCOklahoma: nationwideOKLouisiana: nationwideLAMississippi: nationwideMSAlabama: nationwideALGeorgia: nationwideGAHawaii: nationwideHITexas: named in the noticeTXFlorida: nationwideFL

FDA distribution: “US Nationwide distribution in the states of CT, DE, IN, MA, NC, NJ, NM, NY, OH, PA, TX, VA, WA.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: Connecticut, Delaware, Indiana, Massachusetts, New Jersey, New Mexico, New York, North Carolina, Ohio, Pennsylvania, Texas, Virginia, Washington.

Recall timeline

FDA published this recall 51 days after the firm started it.

1. Firm started the recall: Jun 8, 202612. FDA classified it Class II: Jul 22, 202623. Published in FDA’s Enforcement Report: Jul 29, 202632026
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2774-2026 (June 8, 2026) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2774-2026 to see it in context.
  • Plain-language summary written by qwen3:4b on August 18, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA