Ferno-Washington recalls Power F2 Powered Fastening System (Product No. 0567091) due to potential roller wear and rail deformation that may cause transport delays for patients.
Product may experience accelerated roller wear and potential deformation due to interference between the roller assembly and rail. In the worst-case condition, rail mechanism may become lodged during operation and movement of the stretcher along the rail may no longer be possible through either powered or manual operation. May cause delays in transport of patient to healthcare facility.
Ferno-Washington is recalling its Power F2 Powered Fastening System (Product No. 0567091) because it may experience accelerated roller wear and deformation, which could prevent proper movement of the stretcher during patient transport. This risk could delay critical healthcare delivery.
The roller assembly may interfere with the rail, causing deformation that locks the rail mechanism. This could prevent both powered and manual movement of the stretcher, potentially delaying patient transport to healthcare facilities.
Professional ambulance crews using this system for patient transport are most at risk. The recall affects specific units sold to healthcare facilities.
Check for serial numbers listed in the recall notice: 25C227687 through 26C236253. The product is a powered fastening system for professional ambulance use only.
Verify if your unit has a serial number listed in the recall notice (25C227687 to 26C236253).
The recall states the product may experience accelerated roller wear and deformation, which could prevent movement of the stretcher. Users should check if their unit is affected before continuing use.
The recall does not specify a remedy, so users should check if their unit has a serial number in the list provided and contact Ferno-Washington for further guidance.
If the rail mechanism becomes lodged, movement of the stretcher may not be possible through powered or manual operation, potentially delaying patient transport to healthcare facilities.
We’ve drafted a message you can send Ferno-Washington to request your refund or repair — edit it as you like.
Subject: Recall Z-2760-2026 — Ferno-Washington Ferno-Washington Hello, I own a Ferno-Washington that is covered by recall Z-2760-2026 from Ferno-Washington. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.