ICU Medical Inc. recalled Hotline Fluid Warming Sets on June 24, 2026, due to leakage risks from inadequate solvent application at the tubing-to-connector bond, according to the FDA. The voluntary Class II recall affects 39,750 units distributed nationwide in the U.S.
How to tell if you have it
- Brand: ICU Medical
- Product: Hotline Fluid Warming Set (single use device)
- List Number: L-70
- UDI-DI: 30695085407007
- Lot Numbers: 4293568, 4293569, 4297510, 4297509, 4326087, 4337724, 4337725, 4346349
- Distribution: US Nationwide
The hazard
ICU Medical identified an increased likelihood of leakage in specific lots due to potentially inadequate solvent application at the tubing to return connector bond. The device is designed to warm blood and intravenous fluids under gravity flow conditions.
Context
GovInfoNow's records show ICU Medical, Inc. has 60 other recalls on record since June 15, 2017. Recent brand recalls include the August 20, 2026 CADD Yellow Medication Cassette Reservoir Recall for leakage risk and the May 12, 2026 IV Sets recall regarding black specks in drip chambers. Other recent recalls of similar products include Medline Medical Kits Recall: Particulate in Fluid Path Risk (March 24, 2026) and Medline medical procedure kits with Strykeflow 2 irrigation leak risk (December 16, 2025).