Skip to content
Lower risk Medical device recall · FDA · Published August 4, 2026

VERIQA RT MonteCarlo 3D v3 Software Error May Cause Incorrect Dose Delivery

VERIQA RT MonteCarlo 3D v3 software version 3.1.18 may not clearly display bolus in treatment plans, risking incorrect radiation dose delivery to tissues. Affected units include model S070031.520.

🇺🇸 FDA recall #Z-3174-2026Medical deviceVERIQA ›
✎ GovInfoNow plain-language explainer

VERIQA's RT MonteCarlo 3D v3 software version 3.1.18 may not clearly show bolus doses in treatment plans, which could lead to incorrect radiation delivery to tissues. This recall affects specific software versions used in radiation therapy systems. Read more at govinfonow.com/recall/1134635-veriqa-rt-montecarlo-3d-v3-software-error-may-cause-incorrect.

Source Summary by GovInfoNow · govinfonow.com/recall/1134635-veriqa-rt-montecarlo-3d-v3-software-error-may-cause-incorrect · Updated October 4, 2026 · FDA enforcement record: FDA #Z-3174-2026

Cite this page

“VERIQA RT MonteCarlo 3D v3 Software Error May Cause Incorrect Dose Delivery.” GovInfoNow, updated October 4, 2026. https://govinfonow.com/recall/1134635-veriqa-rt-montecarlo-3d-v3-software-error-may-cause-incorrect. Based on FDA recall #Z-3174-2026 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3174-2026%22&limit=1).

The FDA announced on 2026-08-04 that PTW-FREIBURG is recalling the VERIQA RT MonteCarlo 3D v3 software due to a hazard that may lead users to deliver prescribed doses to the wrong tissue depth. The agency issued recall number Z-3174-2026 for the voluntary Class II action.

How to tell if you have it

  • Brand: VERIQA
  • Product Name: RT MonteCarlo 3D v3
  • Model/Catalog Number: S070031.520
  • Software Version: 3.1.18
  • UDI-DI: +EPTWS070031Q5200

The hazard

When a bolus is included in a treatment plan and sent for secondary dose calculation, the software accounts for it but does not display this clearly. This may lead users to incorrectly omit the bolus during treatment and deliver the prescribed dose to the wrong tissue depth.

Context

GovInfoNow's records show PTW-FREIBURG has issued two other recalls since 2023-06-07, including a 2024-02-08 recall for VERIQA Software Version 2.0/2.1 regarding false positive results risk and a 2023-06-07 recall for BeamAdjust Software 2.3.3 involving incorrect calibration results. Other recent recalls of similar products include RayStation Software Version 12A Recalls (2025-11-28), RayStation 2024B Software Recall (2025-11-28), and RayStation 10B Software May Cause Radiation Dose Errors - Recall (2025-11-28).

How to tell if you have this one

Check for software version 3.1.18 on the VERIQA RT MonteCarlo 3D v3 system with model number S070031.520. The UDI-DI code +EPTWS070031Q5200 may also be present.

  • Brand
    VERIQA
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Software Version

    Verify the software version is 3.1.18 on the VERIQA RT MonteCarlo 3D v3 system with model S070031.520.

Where it went and when

Where it was distributed

Distributed nationwide and abroad; the notice names 2 states.

Alaska: nationwideAKMaine: nationwideMEVermont: nationwideVTNew Hampshire: nationwideNHWashington: nationwideWAIdaho: nationwideIDMontana: nationwideMTNorth Dakota: nationwideNDMinnesota: nationwideMNIllinois: nationwideILWisconsin: nationwideWIMichigan: nationwideMINew York: named in the noticeNYRhode Island: nationwideRIMassachusetts: nationwideMAOregon: nationwideORNevada: nationwideNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: nationwideIAIndiana: nationwideINOhio: nationwideOHPennsylvania: nationwidePANew Jersey: nationwideNJConnecticut: nationwideCTCalifornia: nationwideCAUtah: nationwideUTColorado: nationwideCONebraska: nationwideNEMissouri: nationwideMOKentucky: nationwideKYWest Virginia: nationwideWVVirginia: nationwideVAMaryland: nationwideMDDelaware: nationwideDEArizona: nationwideAZNew Mexico: nationwideNMKansas: nationwideKSArkansas: nationwideARTennessee: nationwideTNNorth Carolina: nationwideNCSouth Carolina: nationwideSCDistrict of Columbia: nationwideDCOklahoma: nationwideOKLouisiana: nationwideLAMississippi: nationwideMSAlabama: nationwideALGeorgia: named in the noticeGAHawaii: nationwideHITexas: nationwideTXFlorida: nationwideFL

FDA distribution: “Worldwide - US Nationwide distribution in the states of GA, NY and the countries of Austria, Belgium, Switzerland, Czechia, Germany, Spain, France, Netherlands, Norway, Poland, Portugal, Romania, T¿rkiye, Chile, United Arab Emirates, Saudi…” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: Georgia, New York.

Recall timeline

FDA published this recall 50 days after the firm started it.

1. Firm started the recall: Aug 4, 202612. FDA classified it Class II: Sep 15, 202623. Published in FDA’s Enforcement Report: Sep 23, 202634. Added to GovInfoNow: Sep 29, 202642026
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. Added to GovInfoNow

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-3174-2026 (August 4, 2026) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-3174-2026 to see it in context.
  • Plain-language summary written by qwen3:4b on October 4, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA