Siemens recalls Artis zee MP systems with material no. 10094139 for potential display issues after power save mode, which may require treatment cancellation or system restart.
After the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem occurs, the system cannot be operated normally. It may be necessary to cancel or restart the treatment or transfer the patient to an alternate or another system.
Siemens is recalling Artis zee MP medical imaging systems with material number 10094139. The display may stay black without showing an error message after returning from power save mode, which could disrupt procedures and require treatment cancellation or system restart.
After the display returns from power save mode, it may show no image and stay black without an error message. This can prevent the system from operating normally during procedures, potentially requiring treatment cancellation or patient transfer to another system.
Healthcare providers using Siemens Artis zee MP systems with material number 10094139 for angiography or fluoroscopy procedures are most at risk. Patients undergoing these procedures may be affected if the system malfunctions.
Check for the material number 10094139 on the Artis zee MP system. This model is used for angiography and whole body radiographic/fluoroscopic procedures.
After the display returns from power save mode, verify it shows an image and not a black screen without error messages.
The recall affects the medical imaging systems used during procedures, but patients are not directly at risk unless the system malfunctions during their treatment.
Look for the material number 10094139 on the Artis zee MP system used for angiography or fluoroscopy procedures.
No immediate stop is required, but if the display stays black without an error message after power save, you should cancel the procedure or restart the system.
We’ve drafted a message you can send Siemens to request your refund or repair — edit it as you like.
Subject: Recall Z-2313-2018 — Siemens Siemens Hello, I own a Siemens that is covered by recall Z-2313-2018 from Siemens. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.