Olympus recalls Bronchovidescope BF-1T240 models due to potential loosening of the instrument channel port when non-Olympus accessories are attached. All units are affected.
The attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Olympus is recalling its Bronchovidescope BF-1T240 models because attaching non-Olympus accessories might cause the instrument channel port to loosen. This could lead to safety issues during use.
When non-Olympus accessories are attached to the instrument channel port, the device may apply more force than expected. This can cause the port to loosen, potentially leading to safety risks during medical procedures.
Medical professionals who use the Bronchovidescope BF-1T240 are most at risk. Patients undergoing procedures with this device may also be affected.
Check for the Bronchovidescope BF-1T240 model with any lot code. The product was sold by Olympus Corporation of the Americas.
Ensure no non-Olympus accessories are attached to the instrument channel port.
The recall affects medical professionals using the device, but patients undergoing procedures with this device may be at risk if the device is used improperly.
Check for the Bronchovidescope BF-1T240 model with any lot code. The product was sold by Olympus Corporation of the Americas.
Ensure no non-Olympus accessories are attached to the instrument channel port. Olympus does not provide a specific remedy for this recall.
We’ve drafted a message you can send Olympus to request your refund or repair — edit it as you like.
Subject: Recall Z-3286-2018 — Olympus Olympus Hello, I own a Olympus that is covered by recall Z-3286-2018 from Olympus. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.