Olympus recalls Bronchovideoscope BF-1T40 models due to potential loosening of the instrument channel port when non-Olympus accessories are attached. Affected units include all lot codes. No remedy specified in recall notice.
The attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Olympus is recalling its Bronchovideoscope BF-1T40 models because attaching non-Olympus accessories might cause the instrument channel port to loosen. This could lead to safety issues during medical procedures. The recall affects all units with the specified model and lot codes.
The hazard occurs when non-Olympus accessories are attached to the instrument channel port, which may apply more force than expected. This can cause the port to loosen, potentially leading to safety risks during medical procedures.
Medical professionals using the Bronchovideoscope BF-1T40 for procedures are most at risk. Patients undergoing procedures with this device could be affected if the instrument channel port becomes loose.
Check for the Bronchovideoscope BF-1T40 model with any lot code. The recall applies to all units of this model.
Identify the model and lot code to confirm if your unit is affected.
The recall notice does not specify a remedy.
All Bronchovideoscope BF-1T40 models with any lot code.
Attaching non-Olympus accessories may cause the instrument channel port to loosen, potentially leading to safety issues during medical procedures.
We’ve drafted a message you can send Olympus to request your refund or repair — edit it as you like.
Subject: Recall Z-3312-2018 — Olympus Olympus Hello, I own a Olympus that is covered by recall Z-3312-2018 from Olympus. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.