GE Healthcare recalls Discovery NM 630 nuclear imaging systems with possible detector mounting failure during installation. Affected units have specific model and serial numbers. Contact GE for details.
During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.
GE Healthcare is recalling certain Discovery NM 630 nuclear imaging systems due to a potential issue with detector mounting during installation. This could lead to loss of one detector mounting design redundancy, which might affect system safety and performance.
During installation of certain Nuclear Medicine systems with dismounted detectors, some detector screws may not have been secured correctly. This could lead to the loss of one detector mounting design redundancy, potentially affecting the system's safety and performance.
Healthcare facilities that own or use the Discovery NM 630 nuclear imaging system. Medical professionals and technicians who operate the equipment are most at risk if the mounting issue affects system safety.
Check for specific model numbers and serial numbers listed in the recall notice: 830Y60040, 830X60004, 830X60001, and other combinations ending with H3910AC. These units were sold with specific lot or serial numbers.
Verify your Discovery NM 630 system's model and serial numbers against the list provided in the recall notice.
The affected models include 830Y60040, 830X60004, 830X60001, and other combinations ending with H3910AC as listed in the recall notice.
During installation of certain Nuclear Medicine systems with dismounted detectors, some detector screws may not have been secured correctly, leading to possible loss of one detector mounting design redundancy.
The recall notice does not specify a remedy, so contact GE Healthcare directly for further instructions.
We’ve drafted a message you can send GE Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-1261-2019 — GE Healthcare GE Healthcare Hello, I own a GE Healthcare that is covered by recall Z-1261-2019 from GE Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.