GE Healthcare recalls NM/CT 860 nuclear medicine systems due to potential detector mounting issues during installation. This could affect system stability and safety in medical imaging procedures.
During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.
GE Healthcare is recalling certain NM/CT 860 nuclear medicine imaging systems because some detector screws may not have been secured properly during installation. This could lead to loss of one detector mounting design redundancy, potentially affecting system stability and safety during medical procedures.
During installation, some detector screws might not be secured correctly, which could cause the loss of one detector mounting design redundancy. This might affect the system's stability and safety during medical imaging procedures.
Medical facilities using GE NM/CT 860 systems for nuclear medicine imaging procedures are likely affected. Staff and patients using these systems may be at risk if the detector mounting issue occurs.
Check for the model numbers H3908AD or H3908AE on the system's label. The systems were shipped with specific lot numbers including 860Y62005, 850X61003, 860X62001, 870X64005, and 860Y62004.
Look for model numbers H3908AD or H3908AE on the system's label to confirm if it's affected.
During installation, some detector screws may not have been secured correctly, leading to possible loss of one detector mounting design redundancy.
Check for model numbers H3908AD or H3908AE on the system's label. Affected systems include those shipped with specific lot numbers like 860Y62005, 850X61003, 860X62001, 870X64005, and 860Y62004.
The recall record does not specify a remedy, so no action is required beyond checking if your system is affected.
We’ve drafted a message you can send GE Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-1263-2019 — GE Healthcare GE Healthcare Hello, I own a GE Healthcare that is covered by recall Z-1263-2019 from GE Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.