Thermo Fisher Scientific recalls precision baths with model TSGP15D that may malfunction and overheat due to assembly errors. Affected units include specific serial numbers.
Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.
Thermo Fisher Scientific is recalling precision baths with model number TSGP15D because they may be assembled incorrectly, causing the bath to overheat beyond the user's set temperature. This could lead to unsafe conditions if the bath is used improperly.
The bath may have been assembled incorrectly, causing the temperature of the fluid in the reservoir to rise above the user-selected temperature setting. This malfunction could lead to unsafe conditions if the bath is used without proper monitoring.
Users of the Thermo Scientific General Purpose Dynamic Bath Precision 5 & 1:0L Dual GP Bath model TSGP15D are affected. Laboratory or industrial users who rely on precise temperature control are at higher risk.
Check for model number TSGP15D on the product. Affected units have specific serial numbers: 300246816, 300246817, 300247411, 300247412.
Look for the model number TSGP15D on the product to confirm it is affected.
Contact Thermo Fisher Scientific (Asheville) LLC for further assistance if you have the affected model.
The model number is TSGP15D.
The affected serial numbers are 300246816, 300246817, 300247411, and 300247412.
The product may have been assembled incorrectly, causing the bath to malfunction and overheat beyond the user's set temperature.
We’ve drafted a message you can send Thermo Fisher to request your refund or repair — edit it as you like.
Subject: Recall Z-1871-2019 — Thermo Fisher Thermo Fisher Hello, I own a Thermo Fisher that is covered by recall Z-1871-2019 from Thermo Fisher. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.