Thermo Fisher Scientific recalls 12 specific model TSGP10 baths that may malfunction due to incorrect assembly, risking overheating. Affected units have serial numbers listed in the recall notice.
Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.
Thermo Fisher Scientific is recalling 12 specific 10L precision baths (model TSGP10) that could malfunction due to incorrect assembly, potentially overheating the fluid. This risk could cause burns or other injuries if the bath is used improperly.
The bath may have been assembled incorrectly, causing it to malfunction and raise the temperature of the fluid above the user-selected setting. This overheating could lead to burns or other injuries if the bath is used without proper temperature control.
Users of the Thermo Scientific General Purpose Dynamic Bath Precision 10L GP Bath model TSGP10 are affected. Laboratory technicians and researchers using these baths for experiments are most at risk due to potential overheating.
Check for the model number TSGP10 on the bath. Affected units have specific serial numbers listed in the recall notice: 300246811, 300246812, 300246814, 300246815, 300247403, 300247404, 300247406, 300247407, 300247409, 300247789, 300247791, 300247792.
Look for the model number TSGP10 on the bath to confirm it is affected.
Contact Thermo Fisher Scientific (Asheville) LLC for further instructions on how to address the issue.
The model number is TSGP10.
Affected serial numbers are 300246811, 300246812, 300246814, 300246815, 300247403, 300247404, 300247406, 300247407, 300247409, 300247789, 300247791, and 300247792.
Check for the model number TSGP10 and the specific serial numbers listed in the recall notice.
We’ve drafted a message you can send Thermo Fisher to request your refund or repair — edit it as you like.
Subject: Recall Z-1868-2019 — Thermo Fisher Thermo Fisher Hello, I own a Thermo Fisher that is covered by recall Z-1868-2019 from Thermo Fisher. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.