Zyno Medical recalls 105" administration sets with inverted pinch clamps that may cause blood to be drawn from patients instead of delivering fluids during infusion. Affected units have specific product codes and lot numbers.
Administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary set is used with the primary set. Thus, instead of delivering fluid to the patient, blood may be drawn from the patient if there is space in the drug container to hold extra volume of fluid, under the condition that the reversed pinch clamp is not identified
Zyno Medical is recalling administration sets that can cause blood to be drawn from patients instead of delivering fluids if the pinch clamp is inverted. This risk occurs when the set is used with certain infusion pumps and the reversed clamp isn't identified.
The administration sets may have an inverted pinch clamp that reverses fluid flow. When used with secondary sets and infusion pumps, blood could be drawn from the patient instead of delivering fluids, especially if the reversed clamp isn't identified.
Healthcare professionals using Zyno Medical administration sets with specific product codes and lot numbers are most at risk. Patients receiving infusions through these sets could experience unintended blood removal.
Check for product code B2-70072-F and lot numbers 18046309 or 18026344 on the administration set. These units were sold with specific UPC/GTIN numbers.
Inspect the administration set for an inverted pinch clamp that may reverse fluid flow.
Reach out to Zyno Medical for assistance with identifying affected units or alternative solutions.
The administration sets may have an inverted pinch clamp that causes blood to be drawn from the patient instead of delivering fluids when used with certain infusion pumps.
Product code B2-70072-F and lot numbers 18046309 or 18026344 are affected.
Check for product code B2-70072-F and lot numbers 18046309 or 18026344 on the administration set.
We’ve drafted a message you can send Zyno Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-2059-2019 — Zyno Medical Zyno Medical Hello, I own a Zyno Medical that is covered by recall Z-2059-2019 from Zyno Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.