Zyno Medical recalls 105" Admin Set 20DP with inverted pinch clamp that may cause blood to be drawn from patients instead of fluid delivery. Affected units have specific lot numbers and product codes.
Administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary set is used with the primary set. Thus, instead of delivering fluid to the patient, blood may be drawn from the patient if there is space in the drug container to hold extra volume of fluid, under the condition that the reversed pinch clamp is not identified
Zyno Medical is recalling its 105" Administration Set 20DP due to a risk that an inverted pinch clamp could cause blood to be drawn from patients instead of delivering medication. This could happen if the set is used with certain infusion pumps and the reversed clamp isn't identified.
The administration set may have an inverted pinch clamp that reverses fluid flow. When used with secondary sets and infusion pumps, this could cause blood to be drawn from the patient instead of medication being delivered, especially if the reversed clamp isn't identified.
Healthcare professionals using Zyno Medical's 105" Administration Set 20DP with infusion pumps are most at risk. Patients receiving treatment with this specific set could experience unintended blood draws.
Check for product code B2-70072 and lot numbers 18037128, 18065375, or 19015969. The set is labeled for use with Zyno Medical infusion pumps or gravity feed systems.
Inspect the administration set for an inverted pinch clamp that may reverse fluid flow.
The administration set may have an inverted pinch clamp that causes blood to be drawn from patients instead of delivering medication when used with certain infusion pumps.
Check for product code B2-70072 and lot numbers 18037128, 18065375, or 19015969. If you have the affected set, inspect for an inverted pinch clamp.
The recall record does not specify a remedy or fix for this issue.
We’ve drafted a message you can send Zyno Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-2058-2019 — Zyno Medical Zyno Medical Hello, I own a Zyno Medical that is covered by recall Z-2058-2019 from Zyno Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.