Wilson-Cook recalls HEMO-7 Hemospray Endoscopic Hemostats with cracked handles that may cause carbon dioxide cartridge to exit during use. Affected units from Jan 2017 to Jan 2020 need inspection.
Complaints that the handle and/or activation knob on Hemostat devices have cracked or broken when the device is activated. In some cases, this has caused the carbon dioxide cartridge to exit the handle. Most of these complaints reported that the carbon dioxide cartridge exited the handle with minimal force. However, some complaints reported that the carbon dioxide cartridge exited the handle with the potential to project towards a user or a patient.
Wilson-Cook is recalling HEMO-7 Hemospray Endoscopic Hemostats because their handles or activation knobs can crack when activated. This may cause the carbon dioxide cartridge to exit the handle, potentially projecting toward users or patients.
When the handle or activation knob cracks during use, the carbon dioxide cartridge can exit the handle. In some cases, this cartridge may project toward a user or patient with enough force to cause injury.
Medical professionals using the HEMO-7 Hemospray during procedures are most at risk. Patients receiving procedures with this device could also be affected if the cartridge exits.
Check for HEMO-7 Hemospray Endoscopic Hemostats with order numbers G56572 (HEMO-7) or G21049 (HEMO-10) manufactured between January 16, 2017 and January 15, 2020.
Inspect the handle and activation knob for cracks or breaks before use.
Yes, the handle or activation knob can crack when activated, potentially causing the carbon dioxide cartridge to exit the handle and project toward users or patients.
Check for HEMO-7 Hemospray Endoscopic Hemostats with order numbers G56572 (HEMO-7) or G21049 (HEMO-10) manufactured between January 16, 2017 and January 15, 2020.
Inspect the handle and activation knob for cracks or breaks before use. If cracked, stop using it and contact Wilson-Cook Medical for further instructions.
We’ve drafted a message you can send Wilson-Cook Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-1745-2020 — Wilson-Cook Medical Wilson-Cook Medical Hello, I own a Wilson-Cook Medical that is covered by recall Z-1745-2020 from Wilson-Cook Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.