Wilson-Cook recalls HEMO-10 Hemospray Endoscopic Hemostats with cracked handles that may cause carbon dioxide to exit unexpectedly. Affected units from Jan 2017 to Jan 2020 need evaluation.
Complaints that the handle and/or activation knob on Hemostat devices have cracked or broken when the device is activated. In some cases, this has caused the carbon dioxide cartridge to exit the handle. Most of these complaints reported that the carbon dioxide cartridge exited the handle with minimal force. However, some complaints reported that the carbon dioxide cartridge exited the handle with the potential to project towards a user or a patient.
Wilson-Cook is recalling HEMO-10 Hemospray Endoscopic Hemostats that have cracked handles or activation knobs. This could cause carbon dioxide to exit the handle, potentially projecting toward users or patients during use.
The handle or activation knob can crack when the device is activated. This may cause carbon dioxide to exit the handle, which could potentially project toward users or patients. Most cases involved minimal force, but some reports indicate a risk of projection.
Medical professionals using the HEMO-10 Hemospray during procedures are most at risk. Patients receiving procedures with this device may also be affected.
Check for cracked handles or activation knobs on the HEMO-10 Hemospray. Affected units were manufactured between January 16, 2017 and January 15, 2020 with specific order numbers.
Inspect the handle and activation knob for cracks or breaks.
The device can crack during use, potentially causing carbon dioxide to exit the handle. Most cases involved minimal force, but some reports indicate a risk of projection toward users or patients.
Check the handle and activation knob for cracks. If cracked, stop using it and contact Wilson-Cook Medical for further instructions.
The affected units were manufactured between January 16, 2017 and January 15, 2020.
We’ve drafted a message you can send Wilson-Cook Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-1746-2020 — Wilson-Cook Medical Wilson-Cook Medical Hello, I own a Wilson-Cook Medical that is covered by recall Z-1746-2020 from Wilson-Cook Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.