Siemens recalls Ysio model 10281013 X-ray systems that may process incorrect tube stand positions, causing unexpected movement if users activate movement functions.
The Ysio system may process incorrect internal values for the position of its ceiling tube stand. If in this situation the user activates longitudinal or transversal stand movement (by using the buttons at the tube head or the remote console), the systems controls may (depending on the values being processed) initiate a correction of the tube stand position, resulting in an unexpected movement.
Siemens is recalling Ysio X-ray systems with model number 10281013 due to a potential risk of unexpected tube stand movement when users activate certain functions. This could lead to inaccurate imaging or unintended patient exposure if not corrected.
The system may process incorrect internal values for the ceiling tube stand position. When users activate longitudinal or transversal stand movement through buttons or remote controls, the system might initiate an unexpected correction of the tube stand position, potentially causing inaccurate imaging or unintended patient exposure.
Healthcare facilities using Siemens Ysio X-ray systems with model number 10281013 are most at risk. Patients and staff using these systems during imaging procedures may be affected.
Check for the model number 10281013 on the Ysio X-ray system. The recall includes specific serial numbers listed in the official notice.
Verify the model number on your Ysio X-ray system to confirm if it matches the recalled model 10281013.
If you have the recalled model, watch for unexpected tube stand movement when activating movement functions.
The recalled model number is 10281013.
Check the model number on your Ysio system; the recall includes specific serial numbers listed in the official notice.
Monitor for unexpected tube stand movement when activating movement functions. Contact Siemens for further guidance.
We’ve drafted a message you can send Siemens to request your refund or repair — edit it as you like.
Subject: Recall Z-1501-2020 — Siemens Siemens Hello, I own a Siemens that is covered by recall Z-1501-2020 from Siemens. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.