Abbott recalls specific Afinion 2 power supplies due to electrostatic discharge causing non-functional units. Affected models include catalog numbers 1116553-1116684 and others. Contact Abbott for details.
The analyzer system may have an electrostatic discharge and sparking of the power supply provided with the analyzers resulting in the power supply becoming non-functional.
Abbott is recalling certain power supplies for the Afinion 2 blood analyzer because they can experience electrostatic discharge and sparking, making them non-functional. This affects users who have the listed catalog numbers and could lead to inaccurate test results if the power supply fails.
The power supply units may develop electrostatic discharge and sparking, which can cause them to become non-functional. This means the analyzer might not work properly, potentially leading to inaccurate test results if the power supply fails during use.
Users of the Afinion 2 blood analyzer with the specific catalog numbers listed in the recall. Most at risk are healthcare professionals and lab technicians who rely on these analyzers for testing.
Check the catalog number on the power supply label. Affected units include catalog numbers 1116553, 1116554, 1116556, 1116557, 1116597, 1116598, 1116663, and others listed in the recall. All serial numbers are affected.
Look for the catalog number on the power supply to confirm if it's affected by this recall.
The power supply may experience electrostatic discharge and sparking, causing it to become non-functional.
Power supplies with catalog numbers 1116553, 1116554, 1116556, 1116557, 1116597, 1116598, 1116663, and others listed in the recall are affected.
The recall states all serial numbers are affected, but the remedy is not specified in the record.
We’ve drafted a message you can send Abbott to request your refund or repair — edit it as you like.
Subject: Recall Z-2192-2020 — Abbott Abbott Hello, I own a Abbott that is covered by recall Z-2192-2020 from Abbott. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.