Inpeco recalls FlexLab Centrifuge Module (FLX-202) with specific firmware versions and serial numbers due to potential unbalanced loads that could damage the centrifuge.
The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.
Inpeco is recalling FlexLab Centrifuge Module tubes with specific firmware versions and serial numbers because the loading algorithm might cause unbalanced loads. This could damage the centrifuge and potentially lead to safety issues if not corrected.
The tube loading algorithm in the FlexLab Centrifuge Module may cause the centrifuge to become unbalanced during operation. Unbalanced loads can damage the centrifuge and potentially cause it to malfunction or fail.
Users of the Inpeco FlexLab Centrifuge Module with the specified firmware versions and serial numbers are affected. Lab technicians and researchers using this equipment are most at risk.
Check if your FlexLab Centrifuge Module has the serial numbers listed in the recall notice or firmware versions CM_1-7-0.H86, CM_1-7-1.H86, xCMd_2-5-0.elf, or xCMd_2-5-1-849fad9.elf.
Review your centrifuge's firmware version to see if it matches the recalled versions: CM_1-7-0.H86, CM_1-7-1.H86, xCMd_2-5-0.elf, or xCMd_2-5-1-849fad9.elf.
Confirm your serial number is listed in the provided list of affected serial numbers.
The recall is due to the tube loading algorithm potentially causing unbalanced loads, which could damage the centrifuge.
Firmware versions CM_1-7-0.H86, CM_1-7-1.H86, xCMd_2-5-0.elf, and xCMd_2-5-1-849fad9.elf are affected.
Check your firmware version and serial number against the list provided in the recall notice.
We’ve drafted a message you can send Inpeco to request your refund or repair — edit it as you like.
Subject: Recall Z-1887-2020 — Inpeco Inpeco Hello, I own a Inpeco that is covered by recall Z-1887-2020 from Inpeco. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.