Inpeco recalls Accelerator a3600 Automation Systems with specific modules due to potential sensor delays that could cause safety issues. Affected units include models FLX-270 to FLX-295 with serial numbers ACP.050 and ACP.146.
The sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the response delay can be up to a maximum delay of 2 seconds, instead of 40ms expected by design.
Inpeco is recalling certain Accelerator a3600 Automation Systems used with specific lab modules because sensors might delay safety responses. This could cause a 2-second delay instead of the expected 40ms, potentially leading to safety risks during lab operations.
The system's safety sensors may activate too slowly, causing delays up to 2 seconds instead of the designed 40ms. This delay could lead to safety issues during lab operations if the system is used without proper safeguards.
Lab technicians and healthcare facilities using the Accelerator a3600 system with the listed modules are most at risk. Affected units have specific part numbers and serial numbers.
Check for the specific module part numbers listed: FLX-270, FLX-271, FLX-276, FLX-282, FLX-283, FLX-274, FLX-209, FLX-289, FLX-290, FLX-219, FLX-204, FLX-214, FLX-286, FLX-295. Also look for serial numbers ACP.050 or ACP.146.
Identify the module part numbers and serial numbers to confirm if your unit is affected.
The sensors that activate safety switches may delay response by up to 2 seconds instead of the designed 40ms, potentially causing safety issues during lab operations.
The Accelerator a3600 system with modules including FLX-270, FLX-271, FLX-276, FLX-282, FLX-283, FLX-274, FLX-209, FLX-289, FLX-290, FLX-219, FLX-204, FLX-214, FLX-286, FLX-295, and serial numbers ACP.050 or ACP.146.
The recall does not specify a remedy, so check the product details to confirm if it's affected and contact Inpeco for further instructions.
We’ve drafted a message you can send Inpeco to request your refund or repair — edit it as you like.
Subject: Recall Z-2115-2020 — Inpeco Inpeco Hello, I own a Inpeco that is covered by recall Z-2115-2020 from Inpeco. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.