Carefusion recalls Safe-T-Centesis 8 Fr Catheter Drainage Trays with potential air leaks that could cause patient complications. Affected units include specific batch numbers and model.
The Universal Drainage Set (UDS) may contain valves that leak internally which may result in the introduction of up to 30cc of air into the patient.
Carefusion is recalling Safe-T-Centesis 8 Fr Catheter Drainage Trays that may leak air into patients. This risk could lead to complications like increased pressure in the body. The recall affects specific batch numbers and model.
The drainage tray may have valves that leak internally, allowing up to 30cc of air to enter the patient. This air can increase pressure in the body, potentially causing complications. However, the risk is not immediate or severe under normal use.
Patients using Safe-T-Centesis 8 Fr Catheter Drainage Trays with specific batch numbers are affected. Healthcare providers and patients at risk are those who have the recalled model.
Check for the model number PIG1280T and batch numbers listed: 0001320864, 0001321078, 0001333781, 0001335698, 0001336588, 0001337539, 0001341632, 0001345260.
Look for the model number PIG1280T and batch numbers listed in the recall notice.
The model number is PIG1280T.
Affected batch numbers are 0001320864, 0001321078, 0001333781, 0001335698, 0001336588, 0001337539, 0001341632, 0001345260.
The recall notice does not specify a remedy.
We’ve drafted a message you can send Carefusion to request your refund or repair — edit it as you like.
Subject: Recall Z-2651-2020 — Carefusion Carefusion Hello, I own a Carefusion that is covered by recall Z-2651-2020 from Carefusion. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.