Carefusion recalls Safe-T PLUS Thora-Para 8 Fr Catheter Drainage Trays with potential air leaks that could introduce up to 30cc of air into patients. Affected units include specific batches and catalog numbers.
The Universal Drainage Set (UDS) may contain valves that leak internally which may result in the introduction of up to 30cc of air into the patient.
Carefusion is recalling Safe-T PLUS Thora-Para 8 Fr Catheter Drainage Trays that may leak air into patients. This could cause air buildup in the body, which might lead to complications for patients using medical drainage systems.
The drainage tray may have valves that leak internally, allowing up to 30cc of air to enter the patient's system. This air buildup could potentially cause complications in medical drainage procedures.
Patients using Safe-T PLUS Thora-Para 8 Fr Catheter Drainage Trays with specific batch numbers are affected. Healthcare providers and patients with medical drainage systems are most at risk.
Check for the catalog number TPT1000SP and batch numbers listed in the recall notice. The affected units include specific batch numbers like 0001319836 and 0001320129.
Look for the catalog number TPT1000SP and batch numbers listed in the recall notice.
The drainage tray may leak internally, allowing up to 30cc of air to enter the patient's system, which could cause complications.
Affected batches include 0001319836, 0001320129, 0001321726, 0001321727, 0001322906, 0001323275, 0001324091, 0001324317, 0001333308, 0001333777, 0001334187, 0001335697, 0001337779, 0001337982, 0001341544, 0001342586, 0001344739, 0001346286.
Check the product details for the catalog number and batch numbers. If affected, contact Carefusion for guidance.
We’ve drafted a message you can send Carefusion to request your refund or repair — edit it as you like.
Subject: Recall Z-2656-2020 — Carefusion Carefusion Hello, I own a Carefusion that is covered by recall Z-2656-2020 from Carefusion. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.