GE Healthcare recalls specific GE NM/CT 850 systems with model H3907AD due to loose rotor bearing screws that may release a rotor bearing during use. Affected units have listed serial numbers.
GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.
GE Healthcare is recalling certain GE NM/CT 850 systems with model number H3907AD because loose rotor bearing screws could cause one of the four rotor bearings to release. This poses a risk of potential damage or injury during medical imaging procedures.
Loose rotor bearing screws in the detector could cause one of the four rotor bearings to release during operation. This may lead to equipment malfunction or injury if a bearing dislodges during medical imaging procedures.
Healthcare facilities using the GE NM/CT 850 system with model number H3907AD and serial numbers listed in the recall are affected. Medical staff and patients using these systems are at risk during imaging procedures.
Check if your GE NM/CT 850 system has model number H3907AD and one of the listed serial numbers: 85EY67048, 850Y61018, 85EY67045, 85EY67046, 85EY67047, 850Z61038, 85EY67043, 850Z61042, 85EZ67056, 850Z61043, 850Z61044, 850Z61045, 850Z61041, 850Z61039, 850Z61040, 85EZ67055, 85EZ67053, 85EZ67054, 850Y61019, 85EY67044, 85EY67049, 85EZ67052, 85EY67050, 85EZ67051.
Verify if your GE NM/CT 850 system has model number H3907AD and one of the listed serial numbers.
The model number is H3907AD.
Affected serial numbers include: 85EY67048, 850Y61018, 85EY67045, 85EY67046, 85EY67047, 850Z61038, 85EY67043, 850Z61042, 85EZ67056, 850Z61043, 850Z61044, 850Z61045, 850Z61041, 850Z61039, 850Z61040, 85EZ67055, 85EZ67053, 85EZ67054, 850Y61019, 85EY67044, 85EY67049, 85EZ67052, 85EY67050, 85EZ67051.
The recall record does not specify a remedy or fix for this issue.
We’ve drafted a message you can send GE Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-0853-2021 — GE Healthcare GE Healthcare Hello, I own a GE Healthcare that is covered by recall Z-0853-2021 from GE Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.