GE Healthcare recalls SIGNA Pioneer MRI systems with software versions PX29.1 and PX25.4 (T) that may show missing slices in 3D images, potentially causing anatomical gaps in scans.
Under certain conditions, some slices may be missing which can lead to a gap of anatomy in the 3D Volume images.
GE Healthcare is recalling some SIGNA Pioneer MRI systems that may show missing slices in 3D images, which could create gaps in anatomical scans. This issue affects the software versions PX29.1 and PX2:5.4 (T) and PX28.3 (China only).
The software may cause some slices to be missing in 3D volume images, creating gaps in anatomical structures. This could lead to less accurate scans but does not pose a direct safety risk.
Patients and medical staff using the SIGNA Pioneer MRI systems with specific software versions are affected. The risk is low as the issue only affects image quality, not safety.
Check if your SIGNA Pioneer MRI has software version PX29.1 or PX25.4 (T) in the United States, or PX28.3 in China. The product is a nuclear magnetic resonance imaging system.
Identify your MRI's software version to see if it matches PX29.1 or PX25.4 (T) in the US, or PX28.3 in China.
Software versions PX29.1 and PX25.4 (T) in the US, and PX28.3 in China.
No, this issue only affects image quality and does not pose a safety risk.
Check the software version on your SIGNA Pioneer MRI system; affected versions are PX29.1, PX25.4 (T), or PX28.3 (China).
We’ve drafted a message you can send GE Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-0134-2022 — GE Healthcare GE Healthcare Hello, I own a GE Healthcare that is covered by recall Z-0134-2022 from GE Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.