GE Healthcare recalls Vscan Extend consoles with serial numbers ending in 5693455 due to battery depletion when not properly turned off before removal or computer connection without AC power.
The firm identified that the Vscan Extend device will not power on because of depleted battery, due to the device not being turned off prior to battery removal or when being connected to a computer without AC power.
GE Healthcare is recalling Vscan Extend consoles that won't power on because the battery depletes when the device isn't turned off before battery removal or when connected to a computer without AC power. This issue affects specific models with serial numbers ending in 5693455, and users should check their device to avoid potential failure.
The device fails to power on when the battery is removed without first turning off the device or when connected to a computer without AC power. This depletion can cause the device to become inoperable during use, potentially disrupting medical procedures.
Users of GE Vscan Extend consoles with serial numbers ending in 5693455 are affected. Medical professionals and healthcare facilities using these devices are most at risk due to potential device failure during critical procedures.
Check your Vscan Extend console's serial number (UDI, Model Number) on the device. Affected units have serial numbers ending in 5693455, such as VH1622261Y, VH161664XA, and others listed in the recall notice.
Look for the serial number on your Vscan Extend console, which should end in 5693455 for affected units.
The device may not power on due to battery depletion from improper handling, such as removing the battery without turning off the device or connecting to a computer without AC power.
Check the serial number on your device; affected units have serial numbers ending in 5693455, such as VH1622261Y or VH161664XA.
The recall does not specify a remedy, so contact GE Healthcare directly for guidance on handling affected devices.
We’ve drafted a message you can send GE Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-0518-2022 — GE Healthcare GE Healthcare Hello, I own a GE Healthcare that is covered by recall Z-0518-2022 from GE Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.