GE Healthcare recalls Centricity PACS RA1000 software versions 3.0-3.2 SP8, 4.0-4.0 SP14, 6.0-6.0 SP10.3, and 7.0-7.0 SP0.0.4.7 due to potential inaccurate distance and area measurements when viewing magnified or scaled-down images.
Potential for Distance and Area measurements to display inaccurate measurement values when performed on magnified images, and specific to ZFP, Distance and Area measurements can display inaccurate measurement values when performed on lossy images that are scaled down from their original resolution.
GE Healthcare is recalling certain versions of their Centricity PACS RA1000 software because measurements of distance and area may be inaccurate when images are magnified or scaled down. This could lead to incorrect medical assessments if used in healthcare settings.
The software may display inaccurate distance and area measurements when images are magnified or scaled down from their original resolution. This could lead to incorrect medical assessments if the measurements are used for critical decisions.
Healthcare facilities using the Centricity PACS RA1000 software with specific versions listed in the recall. Medical professionals and imaging staff who rely on these measurements are most at risk.
Check if your Centricity PACS RA1000 software is version 3.0 through 3.2 SP8, 4.0 through 4.0 SP14, 6.0 through 6.0 SP10.3, or 7.0 through 7.0 SP0.0.4.7. The software versions are identified by specific GTIN/DI codes listed in the recall.
Identify your Centricity PACS RA1000 software version by checking the version number in the system settings or documentation.
Software versions 3.0 through 3.2 SP8, 4.0 through 4.0 SP14, 6.0 through 6.0 SP10.3, and 7.0 through 7.0 SP0.0.4.7 are affected.
Inaccurate distance and area measurements could lead to incorrect medical assessments if used in healthcare settings, potentially impacting patient care decisions.
The recall does not mention a specific fix or remedy; users are advised to check their software version and consider updating to a newer version if available.
We’ve drafted a message you can send GE Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-1233-2022 — GE Healthcare GE Healthcare Hello, I own a GE Healthcare that is covered by recall Z-1233-2022 from GE Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.