Medicina recalls 0.5ml Oral Tip Syringes (Code OT005) due to plunger using unapproved polyethylene material. Affected units may not meet safety standards.
Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification
Medicina is recalling 0.5ml Oral Tip Syringes (Code OT0:005) because the plunger uses an unapproved material that doesn't meet safety specifications. This could lead to improper function or safety risks if the syringe is used.
The plunger component uses polyethylene instead of the required polypropylene material. This unapproved material may not meet safety specifications, potentially causing improper function or safety issues during use.
Users of Medicina Oral Tip Syringes (Code OT005) who purchased the product between 2019 and 2020 are affected. Healthcare professionals and individuals using the syringe for medical purposes are most at risk.
Check for the product code OT005 on the syringe packaging. The batch number HM19024 and UDI-DI (01)05060278503417 are also associated with this recall.
Look for the product code OT005 and batch number HM19024 on the syringe packaging.
The recall states the plunger uses unapproved material that doesn't meet safety specifications, so it is not safe to use.
Check for the product code OT005 and batch number HM19024 on the syringe packaging.
The recall does not specify a remedy, so you should contact Medicina Uk Ltd for further instructions.
We’ve drafted a message you can send Medicina to request your refund or repair — edit it as you like.
Subject: Recall Z-0109-2023 — Medicina Medicina Hello, I own a Medicina that is covered by recall Z-0109-2023 from Medicina. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.