Medicina recalls 1ml oral tip syringes with unapproved polyethylene plunger. This may not meet material specifications. Affected units include specific batch numbers and model codes.
Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification
Medicina is recalling 1ml oral tip syringes with a plunger made from unapproved polyethylene material. This could cause the syringe to not meet safety specifications, potentially leading to improper use or injury.
The plunger uses an unapproved material (polyethylene instead of polypropylene) that doesn't meet the required specifications. This could lead to improper function or safety issues during use, though serious injury is not explicitly stated.
Users of Medicina Oral Tip Syringes 1ml (Code: OT01) with specific batch numbers are affected. Healthcare providers and individuals using these syringes for medical purposes are at risk.
Check for the model code OT01 on the syringe. Affected units include batch numbers HM18147, HM19012, HM1:9025, HM19088, HM19164, HM20063, HM21015, and HM21072.
Look for batch numbers HM18147, HM19012, HM19025, HM19088, HM19164, HM20063, HM21015, or HM21072 on the syringe.
The recalled model is Medicina Oral Tip Syringe 1ml with code OT01.
Affected batch numbers include HM18147, HM19012, HM19025, HM19088, HM19164, HM20063, HM21015, and HM21072.
The recall record does not specify a remedy.
We’ve drafted a message you can send Medicina to request your refund or repair — edit it as you like.
Subject: Recall Z-0110-2023 — Medicina Medicina Hello, I own a Medicina that is covered by recall Z-0110-2023 from Medicina. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.