Full Vision recalls specific Trackmaster treadmill models due to rare risk of failing to stop during a fall, potentially causing injury. Affected units have serial numbers starting with FVDC-7585 to FVDC-7737.
This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.
Full Vision is recalling certain Trackmaster treadmill models (TMX428 110V) because a rare electrical issue could cause the treadmill to fail to stop if someone falls. This risk is very uncommon but serious enough to require immediate action.
A rare electrical problem in the treadmill's control board can cause the safety mechanism to fail when the magnet is pulled loose during a fall. This means the treadmill might not stop moving, increasing the risk of injury.
Treadmill owners with serial numbers starting with FVDC-7585 to FVDC-7737 are affected. Users who exercise on these models are most at risk during falls.
Check your treadmill's serial number on the label. Affected units have serial numbers starting with FVDC-7585 through FVDC-7737. The model is Full Vision Trackmaster TMX428 110V.
Look for the serial number on your treadmill's label. Affected units start with FVDC-7585 to FVDC-7737.
If your treadmill has an affected serial number, stop using it immediately until you contact Full Vision for further instructions.
Full Vision identified a rare electrical issue in the control board that could cause the treadmill to fail to stop during a fall, potentially leading to injury.
Check your treadmill's serial number on the label. Affected units have serial numbers starting with FVDC-7585 through FVDC-7737.
Stop using it immediately and contact Full Vision for further instructions. They will provide details on next steps.
We’ve drafted a message you can send Full Vision to request your refund or repair — edit it as you like.
Subject: Recall Z-0364-2023 — Full Vision Full Vision Hello, I own a Full Vision that is covered by recall Z-0364-2023 from Full Vision. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.