Spacelabs Healthcare recalls UVSL Command Module Model 91496 with software versions 2.03.06 and earlier or 2.04 and later. This may cause clinicians to miss critical events due to missing alarms.
Command modules will display the following when parameter processing is suspended, depending on software version: 1) Older versions will have waveforms present with no alarms, 2) Newer versions have waveforms absent with no alarms. This may cause clinicians to miss a potentially catastrophic event.
Spacelabs Healthcare is recalling its Ultraview SL Command Module Model 91496 with specific software versions. The recall is because older software versions show waveforms without alarms, and newer versions show no waveforms at all, which could lead clinicians to miss serious medical events.
The command module may display waveforms without alarms in older software versions, and show no waveforms at all in newer versions. This means clinicians might not detect life-threatening events that require immediate action.
Healthcare providers using the Ultraview SL Command Module Model 91496 with software versions 2.03.06 and earlier or 2.04 and later are most at risk. Clinicians who rely on this device for patient monitoring could miss critical medical events.
Check if your Ultraview SL Command Module Model 91496 has software version 2.03.06 or earlier, or version 2.04 or later. These specific software versions are part of the recall.
Identify the software version on your Ultraview SL Command Module Model 91496 to determine if it is affected by the recall.
The recall is because older software versions show waveforms without alarms, and newer versions show no waveforms at all, which could lead clinicians to miss critical medical events.
Software versions 2.03.06 and earlier, and software versions 2.04 and later.
The recall record does not specify any remedy actions for affected units.
We’ve drafted a message you can send Spacelabs Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-1503-2023 — Spacelabs Healthcare Spacelabs Healthcare Hello, I own a Spacelabs Healthcare that is covered by recall Z-1503-2023 from Spacelabs Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.