Skytron recalls GS70 Salus Surgical Tables with specific serial numbers due to potential continued movement after button release in Bluetooth or wired mode.
When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause was thought to be confined to Bluetooth/wireless communications between the control pendant and the surgical table. However, during investigation, Mizuho discovered that the failure could be replicated even in wired mode if communication between the pendant and table were somehow interrupted. The root cause appears to be: if communication between the controller and table is interrupted while a button is pressed, the table does not know if/when a button is released. This occurs in wireless or wired mode.
Skytron is recalling certain GS70 Salus Surgical Tables because the table may continue moving even after releasing the control button. This could cause unintended movement during surgery, potentially leading to injury.
If communication between the control pendant and surgical table is interrupted while a button is pressed, the table doesn't know when the button is released. This can cause the table to continue moving even after the user stops pressing the button, which may lead to unintended movement during surgery.
Surgical teams using the GS70 Salus Surgical Tables with the listed serial numbers are most at risk. The hazard affects the table's control system during surgical procedures.
Check the serial number on the table's label. The affected models have specific serial numbers listed in the recall notice, including SUC-20207-0006 through SUC-2022X-0100.
Look for the serial number on the table's label to confirm if it's in the list of affected units.
The table may continue moving after releasing the control button due to communication issues between the control pendant and table, which can happen in both wireless and wired modes.
Check the serial number on the table's label. Affected units have specific serial numbers listed in the recall notice, including SUC-20207-0006 through SUC-2022X-0100.
Check the serial number to confirm if it's in the list of affected units. The recall does not include a remedy, so users should monitor the situation and contact Skytron for further guidance.
We’ve drafted a message you can send Skytron to request your refund or repair — edit it as you like.
Subject: Recall Z-2469-2023 — Skytron Skytron Hello, I own a Skytron that is covered by recall Z-2469-2023 from Skytron. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.