LFS Flat screen support arms for Aurora, Aurora 2, Aurora 3, and other series may have a broken stopper block causing rotation beyond safe limits. This could lead to spring shaft breakage. Affected units include specific model numbers and serial ranges.
The stopper block on the LFS bracket could be broken or missing. Without this stopper block, the flat screen attached to the LFS flat screen system can rotate beyond the original angle limit and it could cause excessive force to be given to the spring shaft of the LFS bracket. Due to the excessive force given to the spring shaft, the spring shaft could break and shear off.
LFS Flat screen support arm systems without surgical lamps for Aurora, Aurora 2, Aurora 3, and other series are recalled because the stopper block might be broken or missing. This can cause the screen to rotate too far, putting excessive force on the spring shaft and potentially breaking it.
The stopper block on the LFS bracket could be broken or missing. Without this block, the flat screen can rotate beyond its safe angle limit, creating excessive force on the spring shaft. This force may cause the spring shaft to break and shear off, potentially causing injury.
Users of LFS flat screen support arms for Aurora, Aurora 2, Aurora 3, and other series models are affected. Those who use these arms for medical equipment are at higher risk due to potential spring shaft failure.
Check if your LFS flat screen support arm has a model number or serial number listed in the provided ranges (e.g., RA000006 to RA001386, BM003066 to BM029325). These are specific to the Aurora, Aurora 2, Aurora 3, and other series.
Inspect the LFS bracket for a broken or missing stopper block. If found, stop using the arm immediately.
If the stopper block is broken or missing, stop using the LFS flat screen support arm immediately to prevent potential spring shaft breakage.
The recall affects LFS flat screen support arms for Aurora, Aurora 2, Aurora 3, Aurora 4, Aurora Astro, Stellar XL, and Stellar series. Specific model numbers and serial ranges are listed in the record.
Check if your LFS arm has a model number or serial number in the provided ranges (e.g., RA000006 to RA001386, BM003066 to BM029325).
If the stopper block is broken or missing, stop using the LFS arm immediately to prevent potential spring shaft breakage.
We’ve drafted a message you can send DKK Dai-Ichi Shomei to request your refund or repair — edit it as you like.
Subject: Recall Z-0072-2024 — DKK Dai-Ichi Shomei DKK Dai-Ichi Shomei Hello, I own a DKK Dai-Ichi Shomei that is covered by recall Z-0072-2024 from DKK Dai-Ichi Shomei. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.