Olympus recalls CYF-VHR Cysto-Nephro Videoscopes due to updated cleaning instructions removing reusable brushes. Affected units require checking serial numbers for recall.
Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
Olympus is recalling certain cysto-nephro videoscopes because their cleaning instructions no longer include reusable brushes. This change means users must use single-use brushes instead of the previously recommended reusable ones.
The recall involves updating the cleaning instructions to remove reusable brushes from the manual. Single-use brushes remain compatible, so the hazard is about cleaning method changes rather than physical danger.
Medical professionals using Olympus CYF-VHR cysto-nephro videoscopes with serial numbers listed in the recall. Those who use reusable cleaning brushes are most at risk.
Check the serial numbers on the videoscope: 04953170310522 or 04953170411298. The product is sold under the brand Olympus.
Review the current cleaning instructions for the videoscope to confirm if reusable brushes are still recommended.
The recall states reusable brushes have been removed from the instructions, so they are no longer recommended. Single-use brushes remain compatible.
Check the serial numbers listed: 04953170310522 or 04953170411298. The recall does not specify a remedy.
The recall involves updated cleaning instructions, not a physical hazard. There is no risk of injury mentioned.
We’ve drafted a message you can send Olympus to request your refund or repair — edit it as you like.
Subject: Recall Z-0690-2024 — Olympus Olympus Hello, I own a Olympus that is covered by recall Z-0690-2024 from Olympus. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.