Olympus recalls Model CYF-5R Cystonephrofiberscope due to updated cleaning instructions removing reusable brushes. Affected units may use single-use brushes only.
Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
Olympus is recalling Model CYF- 5R Cystonephrofiberscopes because the updated instructions no longer mention reusable cleaning brushes. This means users should only use single-use brushes as per the new guidelines.
The hazard involves the removal of reusable cleaning brushes from the instructions. This change means users must now use single-use brushes only, which could lead to inadequate cleaning if not followed properly.
Users of the Olympus Model CYF-5R Cystonephrofiberscope are affected. Those who previously used reusable cleaning brushes may be at risk of improper cleaning.
Check for the model number CYF-5R on the device. The recall includes specific serial numbers listed in the UDI-DI: 04953170328817 04953170411182.
Review the updated instructions for use to confirm single-use brushes are the only option.
The recall instructions have removed reusable brushes from the manual. Only single-use brushes are compatible with the Model CYF-5R.
Check the serial numbers listed in the UDI-DI: 04953170328817 04953170411182 to confirm if your device is affected.
The recall does not require returning the device. Users should follow the updated cleaning instructions with single-use brushes only.
We’ve drafted a message you can send Olympus to request your refund or repair — edit it as you like.
Subject: Recall Z-0685-2024 — Olympus Olympus Hello, I own a Olympus that is covered by recall Z-0685-2024 from Olympus. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.