GE Healthcare recalls specific models of CARESCAPE Canvas 1000 monitors that may lose monitoring during power outages, potentially delaying patient condition recognition. Affected units have unique model numbers and serial numbers listed in the recall.
GE HealthCare has become aware of the potential for loss of monitoring on CARESCAPE Canvas 1000, CARESCAPE Canvas Smart Display, and certain versions of CARESCAPE B850 and CARESCAPE B650 monitors if, following the replacement of the CPU timekeeper battery, a mains power loss occurs during active monitoring. In the unlikely event that this occurs, it can result in the displays or monitors not turning on, after the mains power is restored. This can result in a delay in recognizing changes in patient condition until a replacement monitor is connected.
GE Healthcare is recalling certain CARESCAPE Canvas 1000 monitors that may lose monitoring if a power outage occurs after replacing the CPU timekeeper battery. This could delay recognizing changes in patient condition until a replacement monitor is connected.
If the CPU timekeeper battery is replaced and a power outage occurs while the monitor is actively monitoring, the display may not turn on after power is restored. This could delay recognizing changes in patient condition until a replacement monitor is connected.
Healthcare facilities using the affected CARESCAPE Canvas 1000 monitors. Patients at risk if the monitor fails to display critical information during a power outage.
Check if your monitor has model numbers starting with 5865770 followed by specific suffixes (e.g., 5865770, 5865770-01040169) or serial numbers listed in the recall details.
Identify your monitor's model number, which starts with 5865770 followed by specific suffixes listed in the recall.
Reach out to GE Healthcare for verification and instructions on whether your monitor is affected.
The affected models start with 5865770 followed by specific suffixes listed in the recall, such as 5865770, 5865770-01040169, and others.
Check your monitor's model number against the list of affected models provided in the recall, which includes specific suffixes and serial numbers.
Contact GE Healthcare to verify your monitor's status and follow their instructions for potential remedies.
We’ve drafted a message you can send GE Healthcare Finland to request your refund or repair — edit it as you like.
Subject: Recall Z-2168-2024 — GE Healthcare Finland GE Healthcare Finland Hello, I own a GE Healthcare Finland that is covered by recall Z-2168-2024 from GE Healthcare Finland. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.