GE Healthcare recalls specific models of CARESCAPE Canvas Smart Display and monitors due to potential loss of patient monitoring during power outages after CPU battery replacement. Affected units may not turn on after power restoration, delaying critical patient condition recognition.
GE HealthCare has become aware of the potential for loss of monitoring on CARESCAPE Canvas 1000, CARESCAPE Canvas Smart Display, and certain versions of CARESCAPE B850 and CARESCAPE B650 monitors if, following the replacement of the CPU timekeeper battery, a mains power loss occurs during active monitoring. In the unlikely event that this occurs, it can result in the displays or monitors not turning on, after the mains power is restored. This can result in a delay in recognizing changes in patient condition until a replacement monitor is connected.
GE Healthcare is recalling specific models of CARESCAPE Canvas Smart Display and certain monitors because a power outage after replacing the CPU battery could cause the display to not turn on when power is restored. This might delay recognizing changes in a patient's condition until a replacement monitor is connected.
If the CPU battery is replaced and then a power outage occurs while the monitor is actively monitoring a patient, the display may not turn on when power is restored. This could delay recognizing changes in the patient's condition until a replacement monitor is connected.
Patients using the recalled monitors for medical purposes are most at risk. The recall affects healthcare facilities that use the specific model numbers listed in the notice.
Check for model numbers a) 5700023, b) 5700023-02, or c) 5939590-102 on the device. Serial numbers listed in the recall notice also identify affected units.
If you have replaced the CPU battery recently, monitor for power outages to prevent potential display failure after restoration.
GE Healthcare has identified a potential risk that after replacing the CPU battery, a power outage during active patient monitoring could cause the display to not turn on when power is restored, delaying recognition of patient condition changes.
No action is needed if you have not replaced the CPU battery recently. The recall applies only to units where the battery replacement has occurred.
Check the model number on your device: a) 5700023, b) 5700023-02, or c) 5939590-102. Affected units also have specific serial numbers listed in the recall notice.
We’ve drafted a message you can send GE Healthcare Finland to request your refund or repair — edit it as you like.
Subject: Recall Z-2169-2024 — GE Healthcare Finland GE Healthcare Finland Hello, I own a GE Healthcare Finland that is covered by recall Z-2169-2024 from GE Healthcare Finland. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.