GE Medical Systems Israel recalls specific MyoSPECT cardiac SPECT systems due to potential service handle loosening that could cause injury to service personnel from falling covers.
There is a potential issue regarding the service handles for the internal lead (Pb) covers in that the handles could loosen or release which could cause a cover to drop potentially resulting in injury to service personnel due to the weight of the cover.
GE Medical Systems Israel is recalling certain MyoSPECT cardiac SPECT systems because service handles for internal lead covers might loosen, potentially causing covers to drop and injure service personnel. This recall affects specific models and serial numbers manufactured up to February 2024.
The service handles for internal lead (Pb) covers could loosen or release, which might cause the covers to drop due to their weight. This poses a risk of injury to service personnel during maintenance or service activities.
Service personnel who work with the recalled MyoSPECT cardiac SPECT systems are most at risk. The recall applies to specific models and serial numbers manufactured by GE Medical Systems Israel up to February 2024.
Check for the MyoSPECT System models H3912AA, H3912B, H3912AB, or H3912BD with serial numbers listed in the recall notice. These units were manufactured on or before February 13, 2024, or February 15, 2024, respectively.
Inspect the service handles for internal lead (Pb) covers to ensure they are secure and not loose.
The recall affects MyoSPECT System models H3912AA, H3912B, and MyoSPECT ES models H3912AB, H3912BD manufactured on or before February 13, 2024, or February 15, 2024, respectively.
The service handles for internal lead (Pb) covers could loosen or release, potentially causing covers to drop and injure service personnel due to their weight.
Inspect the service handles for security. If loose, contact GE Medical Systems Israel for guidance on safe handling.
We’ve drafted a message you can send GE Medical Systems Israel to request your refund or repair — edit it as you like.
Subject: Recall Z-2040-2024 — GE Medical Systems Israel GE Medical Systems Israel Hello, I own a GE Medical Systems Israel that is covered by recall Z-2040-2024 from GE Medical Systems Israel. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.