GE Medical Systems Israel recalls Millennium Myos: dual-head cardiac gamma cameras with serial numbers in distribution due to potential detector fall risk if improperly relocated.
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
GE Medical Systems Israel recalls certain Millennium Myosight cardiac gamma cameras because improper relocation without adequate detector support could cause the detector to fall, leading to life-threatening injury. Affected units have serial numbers in distribution and were sold through their channels.
The camera's detector mounting mechanisms could become compromised if the system is moved without proper support. This might cause the detector to fall, potentially causing serious injury or death.
Healthcare facilities using the Millennium Myosight cardiac gamma cameras are affected. Medical staff and patients using these cameras are at risk if the detector falls.
The Millennium Myosight is a dual-head nuclear medicine gamma camera for cardiac imaging. It has serial numbers in distribution and was sold by GE Medical Systems Israel.
Immediately stop using the Millennium Myosight camera if it has been relocated without adequate detector support.
GE Medical Systems Israel has become aware that certain Millennium Myosight cameras could have been transported or relocated without adequate detector support, which might cause the detector to fall and lead to life-threatening injury.
Improper relocation without adequate detector support could cause the detector mounting mechanisms to fail, potentially resulting in a detector fall that causes serious injury.
The Millennium Myosight is a dual-head nuclear medicine gamma camera for cardiac imaging with serial numbers in distribution. Check if your camera has a serial number listed in the distribution records.
We’ve drafted a message you can send GE Medical Systems Israel to request your refund or repair — edit it as you like.
Subject: Recall Z-0298-2026 — GE Medical Systems Israel GE Medical Systems Israel Hello, I own a GE Medical Systems Israel that is covered by recall Z-0298-2026 from GE Medical Systems Israel. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.