Philips recalls Ingenia Elition X MRI systems due to potential loose connections causing fire risk in hospital technical rooms. Affected units have specific model and serial numbers.
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
Philips is recalling Ingenia Elition X MRI systems because a loose connection in the power unit could cause smoke or fire in hospital technical rooms. This hazard affects medical facilities using these systems.
The g-MDU (global Mains Distribution Unit) L3 terminal connection can become loose, creating a hotspot that may lead to smoke or fire alarms in the hospital's technical room. This could disrupt critical medical operations if not addressed.
Hospitals and medical facilities that own or use the Ingenia Elition X MRI systems are most at risk. The recall applies to specific model numbers and serial numbers.
Check for the model number 781358 and serial number starting with 45030 on the device. The recall applies to systems sold in the U.S. with these identifiers.
Identify the model number (781358) and serial number (starting with 45030) on the device.
The g-MDU L3 terminal connection may become loose, creating a hotspot that could cause smoke or fire alarms in hospital technical rooms.
The recall does not specify a remedy, so affected facilities should contact Philips for guidance.
Check for model number 781358 and serial number starting with 45030 on the device.
We’ve drafted a message you can send Philips to request your refund or repair — edit it as you like.
Subject: Recall Z-1852-2024 — Philips Philips Hello, I own a Philips that is covered by recall Z-1852-2024 from Philips. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.