Zyno Medical recalls Z800, Z800F, Z800W, and Z800WF infusion pumps with faulty power supply modules that may cause battery drain when unplugged from AC power. Affected units have specific serial numbers and UDI-DI codes.
Zyno Medical has found that certain Z800, Z800F, Z800W, and Z800WF pumps have faulty power supply modules that will not provide a charge to the rechargeable battery within the pump. This may lead to the battery having insufficient or no charge available to the pump when it is unplugged from the AC mains power.
Zyno Medical is recalling several models of infusion pumps because their power supply modules may not charge the battery properly when unplugged from AC power. This could lead to the pump not functioning correctly if the battery runs out of charge.
The pumps have faulty power supply modules that fail to provide sufficient charge to the rechargeable battery when disconnected from AC power. This means the battery may become depleted, causing the pump to stop working properly without a power source.
Patients using Zyno Medical Z800, Z800F, Z800W, or Z800WF infusion pumps with specific serial numbers or UDI-DI codes are affected. Those with pumps that have the listed serial numbers or UDI-DI codes are most at risk.
Check the pump's UDI-DI code on the label (Z800: 008143771020006, Z800F: 00814371020013, etc.) or serial number (e.g., Z-800: 905283). The affected models include Z800, Z800F, Z800W, and Z800WF.
Look for the UDI-DI code on the pump label to confirm if it matches the listed codes for affected models.
Reach out to Zyno Medical for assistance with the recall process if you have the affected pump.
Z800, Z800F, Z800W, and Z800WF infusion pumps with specific UDI-DI codes and serial numbers listed in the recall notice.
Check the pump's UDI-DI code (e.g., Z800: 008143771020006) or serial number against the list provided in the recall notice.
Contact Zyno Medical directly to confirm if your pump is affected and follow their instructions for the recall process.
We’ve drafted a message you can send Zyno Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-2263-2024 — Zyno Medical Zyno Medical Hello, I own a Zyno Medical that is covered by recall Z-2263-2024 from Zyno Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.