Baxter Healthcare recalls specific infusion pumps with battery codes 35083, 35162, 35195, 35724, 35223, and 36010 due to battery performance issues from prolonged storage.
Battery performance issues related to prolonged storage.
Baxter Healthcare is recalling certain infusion pumps that may have battery performance issues if stored for long periods. This could lead to improper function of the pump, which is critical for medical use.
The battery in these pumps may not perform correctly after prolonged storage, potentially causing the pump to malfunction. This could affect the delivery of medication or fluids to patients, which is critical for medical treatment.
Patients using Baxter infusion pumps with specific product or battery codes listed in the recall are affected. Healthcare providers and patients who use these pumps for medical treatments are at risk if the pump malfunctions due to battery issues.
Check the product code on the pump (35700ABB, 35700BAX, 35700BAX2, 3570009) or battery code (35083, 35162, 35195, 35724, 35223, 36010) to see if it matches the recalled items.
Look for product codes 35700ABB, 35700BAX, 35700BAX2, or 3570009 on the pump.
Check for battery codes 35083, 35162, 35195, 35724, 35223, or 36010 on the battery.
Product codes 35700ABB, 35700BAX, 35700BAX2, and 3570009 are affected.
Battery codes 35083, 35162, 35195, 35724, 35223, and 36010 are affected.
No, the recall does not require immediate stop use; however, you should check the product and battery codes to ensure it is affected.
We’ve drafted a message you can send Baxter Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-1150-2021 — Baxter Healthcare Baxter Healthcare Hello, I own a Baxter Healthcare that is covered by recall Z-1150-2021 from Baxter Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.