Recalled Hillrom Welch Allyn XScribe Cardiac Stress Testing System software versions V6.0.0 and V6.3.2 may assign duplicate Unique Identifiers (UIDs), causing patient identification mismatches with physiological data in PACS systems.
There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.
Baxter Healthcare Corporation is recalling specific versions of their Hillrom Welch Allyn XScribe Cardiac Stress Testing System software due to a potential risk of duplicate patient identifiers. If the system receiving DICOM files relies solely on these unique identifiers, it could lead to incorrect patient data matching.
The software may assign duplicate Unique Identifiers (UIDs) to patient exams. If the Picture Archiving and Communication System (PACS) relies only on these UIDs to accept exams, it could mismatch a patient's identification with their physiological data, potentially leading to incorrect medical records.
Healthcare providers using the XScribe Cardiac Stress Testing System with software versions V6.0.0 or V6.3.2 are most at risk. Patients using these systems for cardiac stress testing may experience incorrect identification if the system's unique identifiers are duplicated.
Check for software versions V6.0.0 or V6.3.2 on the XScribe Cardiac Stress Testing System. The recall includes specific UDI/DI numbers and serial numbers listed in the official notice.
Verify the software version on your XScribe Cardiac Stress Testing System to see if it is V6.0.0 or V6.3.2.
If your software version is affected, contact Baxter Healthcare Corporation for further instructions.
The software may assign duplicate Unique Identifiers (UIDs) to patient exams, which could cause patient identification mismatches with physiological data if the Picture Archiving and Communication System (PACS) relies solely on these UIDs.
Check the software version on your XScribe Cardiac Stress Testing System. If it is V6.0.0 or V6.3.2, your system is affected.
Contact Baxter Healthcare Corporation for further instructions if your software version is V6.0.0 or V6.3.2.
We’ve drafted a message you can send Hillrom Welch Allyn to request your refund or repair — edit it as you like.
Subject: Recall Z-0325-2025 — Hillrom Welch Allyn Hillrom Welch Allyn Hello, I own a Hillrom Welch Allyn that is covered by recall Z-0325-2025 from Hillrom Welch Allyn. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.