Noah Medical recalls Galaxy System (GAL-001) devices with serial numbers CRB-1502 to CRB-1554 due to potential power loss during procedures, which could cause interrupted surgery and lung injury.
Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is on mains or internal battery power; if during procedure, could cause interrupted/aborted procedure, may necessitate manual removal, which could lead to lung injury/pneumothorax.
Noah Medical is recalling Galaxy System (GAL-001) devices with specific serial numbers because a defective component may cause power loss during medical procedures. This could lead to interrupted surgeries and potentially lung injuries if the procedure is stopped abruptly.
The system's power supply may have a defective component that causes sudden power loss during use, either from mains power or battery. If this happens mid-procedure, it could interrupt the surgery and require manual removal of the device, potentially causing lung injury or pneumothorax.
Medical professionals using the Galaxy System (GAL-001) for bronchoscopy procedures are most at risk. Patients undergoing these procedures with affected devices could face complications.
Check the serial number on the device, which is listed as CRB-1502 to CRB-1554. The product is sold under the brand name Noah Medical and model GAL-001.
Look for the serial number on the device, which should be one of the listed CRB-1502 to CRB-1554.
Reach out to Noah Medical for further instructions on how to address the issue.
The recalled device is the Galaxy System (GAL-001).
Power loss during a procedure could cause an interrupted or aborted surgery, requiring manual removal of the device, which might lead to lung injury or pneumothorax.
Check the serial number on your device; affected units have serial numbers from CRB-1502 to CRB-1554.
We’ve drafted a message you can send Noah Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-0693-2025 — Noah Medical Noah Medical Hello, I own a Noah Medical that is covered by recall Z-0693-2025 from Noah Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.