GE Healthcare recalls Carestation 600/700 anesthesia systems with specific power management boards that may shut down unexpectedly during AC power loss, causing temporary ventilation interruptions. Affected units were sold between June 2025 and October 2025.
Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.
GE Healthcare is recalling certain anesthesia systems that may shut down unexpectedly if AC power is lost, causing temporary interruptions in mechanical and manual ventilation. This could affect patient safety during surgery if the system doesn't automatically switch to battery power.
If AC power is interrupted, the anesthesia system may not switch to battery power automatically. This can cause temporary disruptions in mechanical ventilation, manual ventilation, and delivery of anesthetic agents. After rebooting, the system won't return to previous settings, potentially affecting patient safety.
Anesthesia providers using Carestation 600 or 700 Series systems with the specific power management board (2076139-001-S) are affected. Patients undergoing surgery are at risk if the system shuts down unexpectedly during power loss.
Check for the power management board model 2076139-001-S in your Carestation 600 or 700 Series anesthesia system. Units were distributed between June 2, 2025 and October 21, 2025.
Identify if your Carestation 600 or 700 Series system has the power management board model 2076139-001-S.
If AC power is lost, the system may not automatically switch to battery power and could shut down, causing temporary interruptions in ventilation and anesthetic delivery.
Check for the power management board model 2076139-001-S in your Carestation 600 or 700 Series system. Affected units were distributed between June 2, 2025 and October 21, 2025.
The recall requires checking the power board model. There is no immediate action needed beyond identifying the affected model.
We’ve drafted a message you can send GE Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-0915-2026 — GE Healthcare GE Healthcare Hello, I own a GE Healthcare that is covered by recall Z-0915-2026 from GE Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.