Wilson-Cook Medical recalls HEMO-7-EU Hemospray Endoscopic Hemostats with specific lots where the activation knob may crack during use, posing a risk of injury. Affected units have unique lot numbers and expiration dates.
Specific lots of product have been manufactured with nonconforming raw materials resulting in the red activation knob cracking or breaking at the activation knob internal threading while the device is activated.
Wilson-Cook Medical is recalling certain HEMO-7-EU Hemospray Endoscopic Hemostats due to a risk that the activation knob could crack while in use. This could lead to device failure during medical procedures, potentially causing injury to patients or healthcare providers.
The activation knob in specific lots can crack or break at the internal threading when the device is activated. This could cause the device to malfunction during medical procedures, potentially leading to injury or complications.
Healthcare professionals using the HEMO-7-EU Hemospray Endoscopic Hemostat for medical procedures are most at risk. Patients receiving care during these procedures may also be affected if the device fails.
Check the product's lot numbers: W4844245 (expires 6/10/2027), W4847501 (expires 6/20/2027), or W4851589 (expires 7/8/2027). The device is a 2.8 cm sterile endoscopic hemostat with UDI-DI 00827002246638.
Look for lot numbers W4844245, W4847501, or W4851589 on the product packaging or device label.
Reach out to Wilson-Cook Medical for further assistance or to confirm if your device is affected.
The activation knob in specific lots can crack or break at the internal threading when the device is activated, potentially causing device failure during medical procedures.
Lot numbers W4844245 (expires 6/10/2027), W4847501 (expires 6/20/2027), and W4851589 (expires 7/8/2027).
Check the lot number on the device label; affected units have specific lot numbers: W4844245, W4847501, or W4851589.
We’ve drafted a message you can send Wilson-Cook Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-0582-2025 — Wilson-Cook Medical Wilson-Cook Medical Hello, I own a Wilson-Cook Medical that is covered by recall Z-0582-2025 from Wilson-Cook Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.