Leica Biosystems recalls CM1950 cryostats with firmware V3.01 (model 1491950C4US) due to potential loss of patient tissue. Affected units include specific serial numbers and model numbers.
Firm became aware of software issues related to most recent firmware update which could result in loss of patient tissue.
Leica Biosystems is recalling CM1950 cryostats with firmware version V3.01 (model : 1491950C4US) because software issues could cause loss of patient tissue. This affects specific units with known serial numbers and model numbers.
Software issues in the most recent firmware update could lead to loss of patient tissue during use. This is a potential risk that may affect the accuracy or safety of tissue samples.
Users of Leica Biosystems CM1950 cryostats with firmware version V3.01 (model 1491950C4US) and specific serial numbers are affected. Medical facilities using these units are most at risk.
Check for the model number 1491950C4US and firmware version V3.01 on your CM1950 cryostat. Affected units have specific serial numbers: 12835 and 12854.
Verify your cryostat's firmware version by checking the model number and serial number on the device.
Firmware version V3.01 is affected for model 1491950C4US.
Serial numbers 12835 and 12854 are affected.
The recall does not specify a stop-use action; check firmware version to determine if your unit is affected.
We’ve drafted a message you can send Leica Biosystems to request your refund or repair — edit it as you like.
Subject: Recall Z-1247-2025 — Leica Biosystems Leica Biosystems Hello, I own a Leica Biosystems that is covered by recall Z-1247-2025 from Leica Biosystems. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.