Leica Biosystems recalls CM1900 cryostats with specific material numbers due to risk of serious injury from flammable freezing sprays used in chamber. Check serial numbers for affected units.
As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
Leica Biosystems is recalling CM1900 cryostats that may have been used with flammable freezing sprays. This poses a serious injury risk if the sprays ignite inside the chamber, which can happen if the cryostat label didn't warn about this before 2019.
The cryostat chamber can ignite flammable freezing sprays if they are used inside, potentially causing serious injuries. The label did not warn about this risk before 2019, so users might not have known to avoid it.
Users of Leica Biosystems CM1900 cryostats in labs or medical facilities who use flammable freezing sprays. Those who have used the cryostat without checking for the specific material numbers or serial numbers are most at risk.
Check the material numbers: 14045227382, 14901900100, 14901900700, or 1491900UVU2. Also look for serial numbers listed in the recall notice for affected units.
Identify if your cryostat has material numbers 14045227382, 14901900100, 14901900700, or 1491900UVU2.
Check for serial numbers listed in the recall notice for affected units.
The cryostat chamber can ignite flammable freezing sprays if used inside, potentially causing serious injuries. The label did not warn about this risk before 2019.
Check if your cryostat has material numbers 14045227382, 14901900100, 14901900700, or 1491900UVU2, and look for the specific serial numbers listed in the recall notice.
No. The recall requires checking material numbers and serial numbers to confirm if your unit is affected. There is no immediate stop-use instruction.
We’ve drafted a message you can send Leica Biosystems to request your refund or repair — edit it as you like.
Subject: Recall Z-0088-2025 — Leica Biosystems Leica Biosystems Hello, I own a Leica Biosystems that is covered by recall Z-0088-2025 from Leica Biosystems. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.