GE Healthcare recalls certain Revolution Apex and Revolution CT systems due to potential blue coolant leaks near gantry base that could cause slips and falls. No immediate action needed but monitor for leaks.
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
GE Healthcare is recalling specific models of Revolution CT scanners because coolant fluid (glycol) can leak near the base of the machine, appearing blue. This fluid isn't corrosive or hot but poses a slip and fall risk if not noticed.
The recalled systems may leak blue coolant fluid near the base of the gantry. While the fluid isn't corrosive or hot, it can create a slippery surface if not noticed, potentially causing slips and falls.
Medical facilities using the recalled GE Healthcare Revolution Apex or Revolution CT systems are most at risk. The specific models are identified by unique system IDs and serial numbers listed in the recall notice.
Check for system IDs and serial numbers listed in the recall notice. The affected models include specific system IDs like 907729APEX1, 205608CT, and others with corresponding serial numbers.
Look for visible blue coolant fluid near the base of the gantry. If present, stop using the system immediately.
Reach out to GE Healthcare for assistance with the recall process using the provided system IDs and serial numbers.
No, the coolant fluid is not corrosive and not at a harmful temperature. However, it can cause slips and falls if not noticed.
Check your system ID and serial number against the list provided in the recall notice. Affected models include specific system IDs like 907729APEX1 and 205608CT.
No immediate action is required, but if you see blue fluid near the base of the gantry, stop using the system immediately to avoid slips and falls.
We’ve drafted a message you can send GE Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-1299-2025 — GE Healthcare GE Healthcare Hello, I own a GE Healthcare that is covered by recall Z-1299-2025 from GE Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.