GE Healthcare recalls certain Revolution Apex Plus CT systems due to potential blue coolant fluid leaking near the gantry base, which could cause slips and falls if not noticed. Affected units have specific system IDs and serial numbers.
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
GE Healthcare is recalling specific models of their Revolution Apex Plus CT scanners because blue coolant fluid can leak near the base of the machine. If someone slips on this fluid, it could cause injury. The recall affects certain units with unique system IDs and serial numbers.
The coolant fluid leaks near the base of the gantry and appears blue. While it's not corrosive or hot, it can cause slips and falls if not noticed, leading to potential injuries.
Healthcare facilities using the recalled CT systems are most at risk. Patients and staff could be injured if they slip on the blue coolant fluid that leaks near the machine base.
Check for specific system IDs and serial numbers listed in the recall notice: 907729APEX1, 336832APEX, 973971CTREV, and 513584APEX1. These units were manufactured with the described system IDs and serial numbers.
Look for blue coolant fluid near the base of the gantry. If present, stop using the system immediately.
Reach out to GE Healthcare for guidance on next steps if you have the recalled system.
The fluid is not corrosive or hot, but it can cause slips and falls if not noticed, leading to potential injury.
Check for specific system IDs and serial numbers listed in the recall notice: 907729APEX1, 336832APEX, 973971CTREV, and 513584APEX1.
If you don't see blue fluid, the system may not be affected, but you should still check the system ID and serial number to confirm.
We’ve drafted a message you can send GE Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-1300-2025 — GE Healthcare GE Healthcare Hello, I own a GE Healthcare that is covered by recall Z-1300-2025 from GE Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.